LAN-XI UA-2113 General-purpose Front Panel + Monitor Data acquisition systems
LAN-XI UA-2113 General-purpose Front Panel + Monitor: data acquisition systems by HBK.
This series on the HBK site (opens in a new tab)Models in the series
One column per model: what differs between the models of the series.
| Parameter | UA-2113–060- eli-m-a009eb188f55020a974b45f2 |
|---|---|
| interface_specification | Unknown |
| Protection class (DIN EN 60529) | ip31 |
| Number of measurement channels | 6 |
| quantity_specification | Inputs: Standardized voltage (+/- 10V), IEPE / CCLD; Outputs: Voltage (+/- 10V); Housing: Standalone; Interfaces: None |
| Status | Active |
The industries we serve
Frequently asked questions
Answers to the most common questions about our products and services
Buffered monitor outputs let the same input signal feed both LAN-XI and a second system, such as a recorder.
The panel has 6 SMB input connectors and 6 SMB monitor-output connectors, parallel to those inputs.
An input above 10 V peak clips at the monitor output. The manufacturer therefore advises against monitoring when the module uses its extended 31.6 V peak input range.
The panel supports module 3050-A-060, direct-voltage measurements and CCLD sensors such as accelerometers and microphones.
No. It is a buffered analogue output parallel to the signal input. The second system performs its own acquisition, so check its calibration, sampling and synchronization separately when comparing recordings.
Get a quote or advice
Tell us what you need and an engineer will come back to you.