LAN-XI UA-2114 Dynamic Bridge Front Panel Data acquisition systems
LAN-XI UA-2114 Dynamic Bridge Front Panel: data acquisition systems by HBK.
This series on the HBK site (opens in a new tab)Models in the series
One column per model: what differs between the models of the series.
| Parameter | UA-2114–042- eli-m-1d3bcb9f8ad19899a6ba5794 |
|---|---|
| quantity_specification | Inputs: Standardized voltage (+/- 10V); Outputs: Voltage (+/- 10V); Housing: Standalone; Interfaces: None |
| interface_specification | Unknown |
| Number of measurement channels | 4 |
| Protection class (DIN EN 60529) | ip31 |
| Status | Active |
The industries we serve
Frequently asked questions
Answers to the most common questions about our products and services
The panel is intended for dynamic bridge transducers, such as Kulite LQ-080 and LQ-125, in aircraft and wind-tunnel measurements.
The panel provides ±5 V transducer excitation and 30 dB gain. Enter that gain in the measurement software transducer setup.
DC coupling is possible, but transducer offset matters. An offset above approximately 10 mV moves the Dyn-X input into its upper range and reduces available dynamic range.
UA-2114-030 fits 3052-A-030, UA-2114-060 fits 3050-A-060, and UA-2114-042 fits 3160-A-042 or 3050-A-060. Version 042 adds 2 BNC generator connectors, available with a module providing a generator.
Excitation is shared across channels. The specification states that excessive current drawn on one channel produces an overload indication on all channels. Distinguish a shared excitation fault from signal overload.
Get a quote or advice
Tell us what you need and an engineer will come back to you.