LAN-XI UA-2115 Voltage Injection Calibration Front Panel Data acquisition systems
LAN-XI UA-2115 Voltage Injection Calibration Front Panel: data acquisition systems by HBK.
This series on the HBK site (opens in a new tab)
Models in the series
One column per model: what differs between the models of the series.
| Parameter | UA-2115–042- eli-m-d5d115b7b19a18d80832709c |
|---|---|
| interface_specification | Unknown |
| Number of measurement channels | 4 |
| quantity_specification | Inputs: Standardized voltage (+/- 10V); Outputs: Voltage (+/- 10V); Housing: Standalone; Interfaces: None |
| Protection class (DIN EN 60529) | ip31 |
| Status | Active |
The industries we serve
Frequently asked questions
Answers to the most common questions about our products and services
The required combination is UA-2115-042, module 3160-A-042 and microphone preamplifier 2669-W-004, designed to support voltage injection calibration.
Set polarization voltage in the software to 0 V for VIC or 200 V for CIC, using the specified hardware combination.
The panel has 4 input LEMO connectors with 7 pins and 2 BNC outputs.
A turquoise panel LED indicates VIC. Also check the software calibration mode and the connected preamplifier.
No. Its BNC connections are outputs, while microphone inputs use LEMO connections. Add suitable acquisition channels when increasing microphone count; do not count output connectors as extra inputs.
Get a quote or advice
Tell us what you need and an engineer will come back to you.