capaNCDT CSH/FL/SE Capacitive
capaNCDT CSH/FL/SE: sensors and measurement devices by Micro-Epsilon for capacitive.
Download the datasheet (opens in a new tab) This series on the Micro-Epsilon site (opens in a new tab)
Shared technical specifications for the series
Properties every model in the series has. What differs is in the models table below.
Range across models
The span covered by the models of the series together — not the range of each model.
| Linearity | 0.25–15 µm |
|---|---|
| Measuring range | 0.125–6 mm |
| Measuring window | 0–6 mm |
| Resolution | 0.000075–0.06 µm |
| active_area_published_size | 2.9–13 mm |
| Number of measurement channels | 1–8 |
| Recommended target diameter | 5–30 mm |
| Weight | 0.028–0.037 kg |
| ADC sampling rate | 1000–8500 samples/s |
| Accuracy | 0.25–15 µm |
Models in the series
One column per model: what differs between the models of the series.
| Parameter | CSH-0,25/FL/SE/CRm-1,4 eli-m-fda4d3bb60997465ba47a3d6 | CSH-0,5/FL/SE/CRm-1,4 eli-m-48c626360e2ff13d79684e1b | CSH-1,2/FL/SE/CRm-1,4 eli-m-68423c43fa840df6633a9601 | CSH-1/FL/SE/CRm-1,4 eli-m-12fbde90119268ef231ee665 | CSH-2/FL/SE/CRm-1,4 eli-m-e0054949f43c57e374e8a561 | CSH-3/FL/SE/CRm-1,4 eli-m-9a429c8cb42493763585e361 | CSH-5/FL/SE/CRm-1,4 eli-m-61ef25449da5e0c46b64efa2 |
|---|---|---|---|---|---|---|---|
| Linearity | 0.25 µm | 0.35 µm | 1.2 µm | 0.6 µm | 0.6 µm | 1.5 µm | 15 µm |
| Start of measuring range | 0 mm | 0 mm | 0 mm | 0 mm | 0 mm | 0 mm | 0 mm |
| source_sampling_rate_basis | DT6112 high-speed controller: 20 kHz bandwidth | DL6220: 5 kHz bandwidth; DT6240 digital rate 3.906 kSamples/s | DL6220: 5 kHz bandwidth; DT6240 digital rate 3.906 kSamples/s | DT6110: 1 kHz analog measurement bandwidth | DL6510 with CP6001: 8.5 kHz bandwidth in the official maximum-eight-channel DT6530 base system | DL6510 with CP6001: 8.5 kHz bandwidth in the official maximum-two-channel DT6530 base system | DT6110/IP: IP68 single-channel controller with 1 kHz bandwidth |
| interface_specification | Analog 0-10V | 4-20mA | 4-20mA | Analog 0-10V | Analog 0-10V | Analog 0-10V | 4-20mA |
| Measuring range | 0.125 mm | 0.25 mm | 1.2 mm | 1 mm | 4 mm | 6 mm | 5 mm |
| Interfaces | ethernet | ethercat | ethercat | ethernet | ethercat | profinet | ethernet |
| Protection class (DIN EN 60529) | ip40 | ip40 | ip40 | ip40 | ip40 | ip40 | ip40 |
| Measuring window | 0–0.25 mm | 0–1 mm | 0–0.6 mm | 0–0.5 mm | 0–1 mm | 0–6 mm | 0–2.5 mm |
| Operating temperature | -269–200 °C | -269–200 °C | -269–200 °C | -269–200 °C | -269–200 °C | -269–200 °C | -269–200 °C |
| source_linearity_basis | Official sensor linearity from the current capaNCDT catalog; total measuring-channel linearity also includes the DT6222 + DL6222 controller contribution. | Official sensor linearity from the current capaNCDT catalog; total measuring-channel linearity also includes the DT6110 controller contribution. | Official sensor linearity from the current capaNCDT catalog; total measuring-channel linearity also includes the DT6120/IP controller contribution. | Official sensor linearity from the current capaNCDT catalog; total measuring-channel linearity also includes the DT6220 + DL6230 controller contribution. | Official sensor linearity from the current capaNCDT catalog; total measuring-channel linearity also includes the DT6240 + DL6230 controller contribution. | Official sensor linearity from the current capaNCDT catalog; total measuring-channel linearity also includes the DT6238 + DL6228 controller contribution. | Official sensor linearity from the current capaNCDT catalog; total measuring-channel linearity also includes the DT6238 + DL6228 controller contribution. |
| Resolution | 7.499999999999999e-05 µm | 0.00015 µm | 0.00036 µm | 0.0003 µm | 0.0005999999999999999 µm | 0.05999999999999999 µm | 0.0015 µm |
| quantity_specification | 0-0.25 mm model-level measuring interval | 0-0.5 mm model-level measuring interval | 0-1.2 mm model-level measuring interval | 0-1 mm model-level measuring interval | 0-2 mm model-level measuring interval | 0-3 mm model-level measuring interval | 0-5 mm model-level measuring interval |
| active_area_published_size | 2.9 mm | 4.1 mm | 6.3 mm | 5.7 mm | 8.1 mm | 10 mm | 13 mm |
| Number of measurement channels | 1 | 4 | 2 | 8 | 1 | 4 | 4 |
| Recommended target diameter | 5 mm | 7 mm | 11 mm | 11 mm | 17 mm | 24 mm | 30 mm |
| Storage temperature | -269–200 °C | -269–200 °C | -269–200 °C | -269–200 °C | -269–200 °C | -269–200 °C | -269–200 °C |
| Weight | 0.028 kg | 0.028 kg | 0.03 kg | 0.03 kg | 0.037 kg | 0.037 kg | 0.037 kg |
| Contact | Non-contact | Non-contact | Non-contact | Non-contact | Non-contact | Non-contact | Non-contact |
| ADC sampling rate | 5000 samples/s | 5000 samples/s | 8500 samples/s | 2000 samples/s | 8500 samples/s | 1000 samples/s | 8500 samples/s |
| Accuracy | 0.25 µm | 0.35 µm | 1.2 µm | 0.6 µm | 0.6 µm | 1.5 µm | 15 µm |
| Status | Active | Active | Active | Active | Active | Active | Active |
| Datasheet | PDF — CSH-0,25/FL/SE/CRm-1,4 (opens in a new tab) | PDF — CSH-0,5/FL/SE/CRm-1,4 (opens in a new tab) | PDF — CSH-1,2/FL/SE/CRm-1,4 (opens in a new tab) | PDF — CSH-1/FL/SE/CRm-1,4 (opens in a new tab) | PDF — CSH-2/FL/SE/CRm-1,4 (opens in a new tab) | PDF — CSH-3/FL/SE/CRm-1,4 (opens in a new tab) | PDF — CSH-5/FL/SE/CRm-1,4 (opens in a new tab) |
The industries we serve
Frequently asked questions
Answers to the most common questions about our products and services
CSH/FL/SE uses a flat rather than cylindrical design. It also addresses vacuum, cleanroom and strong-magnetic-field environments.
No. Recommended flat-target size depends on model span and differs from active area. Use the model column, not the smallest family value.
Yes. The table distinguishes front and rear mounting temperature coefficients. Hole count and type also depend on model.
It specifies an integral 1.4 m cable with a type B connector. Other cables, feedthroughs and controller must also match the special environment.
No. Its explicit rating is IP40. Housing material and special-environment suitability do not increase that rating.
Get a quote or advice