confocalDT IFD2410
confocalDT IFD2410: sensors and measurement devices by Micro-Epsilon for distance & displacement sensors.
Download the datasheet (opens in a new tab) This series on the Micro-Epsilon site (opens in a new tab)
Shared technical specifications for the series
Properties every model in the series has. What differs is in the models table below.
Range across models
The span covered by the models of the series together — not the range of each model.
| Protection class (DIN EN 60529) | ip65 |
|---|---|
| Light spot diameter | 0.012–0.024 mm |
| Minimum target thickness | 0.05–0.3 mm |
| Maximum measuring angle | 10–25 deg |
| Start of measuring range | 15–35 mm |
| Output resolution | 18–32 bit |
| Resolution | 0.006–0.25 µm |
| Linearity | 0.25–1.5 µm |
| Measuring range | 1–6 mm |
Models in the series
One column per model: what differs between the models of the series.
| Parameter | IFD2410-1 eli-m-f66806132201ca4612e74361 | IFD2410-1/IE eli-m-4a36c0bcb9e929f457be0850 | IFD2410-3 eli-m-b01c8b9c6d70ab5f63fc17e0 | IFD2410-3/IE eli-m-4dbdc1f43486faee54f34ca4 | IFD2410-6 eli-m-47255602b5dbb60eba71061a | IFD2410-6/IE eli-m-56987f833f4c4dae7883a789 |
|---|---|---|---|---|---|---|
| Protection class (DIN EN 60529) | ip65 | ip64 | ip65 | ip64 | ip64 | ip64 |
| Interfaces | ethernet | ethercat | ethercat | profinet | ethercat | ethercat |
| source_permissible_ambient_illumination_lux | 30,000 lx | 30,000 lx | 30,000 lx | 30,000 lx | 30,000 lx | 30,000 lx |
| Light spot diameter | 0.012 mm | 0.012 mm | 0.018 mm | 0.018 mm | 0.024 mm | 0.024 mm |
| source_target_material | Reflective, diffuse as well as transparent surfaces (e.g. glass) | Reflective, diffuse as well as transparent surfaces (e.g. glass) | Reflective, diffuse as well as transparent surfaces (e.g. glass) | Reflective, diffuse as well as transparent surfaces (e.g. glass) | Reflective, diffuse as well as transparent surfaces (e.g. glass) | Reflective, diffuse as well as transparent surfaces (e.g. glass) |
| Minimum target thickness | 0.05 mm | 0.05 mm | 0.15 mm | 0.15 mm | 0.3 mm | 0.3 mm |
| Number of measurement channels | 1 | Not specified | 1 | Not specified | 1 | Not specified |
| Maximum measuring angle | 25 deg | 25 deg | 19 deg | 19 deg | 10 deg | 10 deg |
| rate_interval | 100–8000 Hz | Not specified | Not specified | Not specified | Not specified | Not specified |
| Start of measuring range | 15 mm | 15 mm | 25 mm | 25 mm | 35 mm | 35 mm |
| Output resolution | 32 bit | Not specified | 32 bit | Not specified | 18 bit | Not specified |
| Resolution | 0.05 µm | 0.006 µm | 0.018 µm | 0.018 µm | 0.25 µm | 0.25 µm |
| source_multi_peak_measurement_layers | 1 layer | 1 layer | 1 layer | 1 layer | 1 layer | 1 layer |
| Linearity | 0.25 µm | Not specified | 0.75 µm | Not specified | 1.5 µm | Not specified |
| source_electrical_connection | 12-pin M12 for supply, Ethernet, RS422 and sync; 17-pin M12 for I/O, analog and encoder; optional 3/6/9/15 m cable extension. | 12-pin M12 for supply, encoder, EtherCAT, PROFINET, EtherNet/IP, RS422 and sync; 17-pin M12 for I/O, analog and encoder; optional 3/6/9/15 m cable extension. | 12-pin M12 for supply, Ethernet, RS422 and sync; 17-pin M12 for I/O, analog and encoder; optional 3/6/9/15 m cable extension. | 12-pin M12 for supply, encoder, EtherCAT, PROFINET, EtherNet/IP, RS422 and sync; 17-pin M12 for I/O, analog and encoder; optional 3/6/9/15 m cable extension. | 12-pin M12 for supply, Ethernet, RS422 and sync; 17-pin M12 for I/O, analog and encoder; optional 3/6/9/15 m cable extension. | 12-pin M12 for supply, encoder, EtherCAT, PROFINET, EtherNet/IP, RS422 and sync; 17-pin M12 for I/O, analog and encoder; optional 3/6/9/15 m cable extension. |
| quantity_specification | 1 mm net measuring range | Not specified | 3 mm net measuring range | Not specified | 6 mm net measuring range | Not specified |
| Storage temperature | -20–70 °C | -20–70 °C | -20–70 °C | -20–70 °C | -20–70 °C | -20–70 °C |
| Power consumption | 5.3 W | 5.3 W | 5.3 W | 5.3 W | 5.3 W | 5.3 W |
| Measuring range | 1 mm | 1 mm | 3 mm | 3 mm | 6 mm | 6 mm |
| Measuring rate | 8000 Hz | Not specified | Not specified | Not specified | Not specified | Not specified |
| Weight | 0.49 kg | 0.49 kg | 0.49 kg | 0.49 kg | 0.49 kg | 0.49 kg |
| Accuracy | 0.2 µm | Not specified | 0.2 µm | Not specified | 0.2 µm | Not specified |
| source_mounting_description | Radial clamping (see accessories for mounting adapter), threaded holes | Radial clamping (see accessories for mounting adapter), threaded holes | Radial clamping (see accessories for mounting adapter), threaded holes | Radial clamping (see accessories for mounting adapter), threaded holes | Radial clamping (see accessories for mounting adapter), threaded holes | Radial clamping (see accessories for mounting adapter), threaded holes |
| source_sampling_rate_basis | IFD2410 integrated controller: up to 25 kHz on the manufacturer product page | Not specified | IFD2410 integrated controller: up to 25 kHz on the manufacturer product page | Not specified | IFD2410 integrated controller: up to 25 kHz on the manufacturer product page | Not specified |
| source_linearity_basis | https://www.micro-epsilon.com/fileadmin/download/products/dat--confocalDT-IFD2410--en.pdf published optical-probe linearity; axial resolution is not substituted for linearity. | Not specified | https://www.micro-epsilon.com/fileadmin/download/products/dat--confocalDT-IFD2410--en.pdf published optical-probe linearity; axial resolution is not substituted for linearity. | Not specified | https://www.micro-epsilon.com/fileadmin/download/products/dat--confocalDT-IFD2410--en.pdf published optical-probe linearity; axial resolution is not substituted for linearity. | Not specified |
| Operating temperature | 5–50 °C | 5–50 °C | 5–50 °C | 5–50 °C | 5–50 °C | 5–50 °C |
| source_numerical_aperture | 0.45 | 0.45 | 0.35 | 0.35 | 0.18 | 0.18 |
| interface_specification | Digital I/O | Not specified | Digital I/O | Not specified | Digital I/O | Not specified |
| Contact | Non-contact | Non-contact | Non-contact | Non-contact | Non-contact | Non-contact |
| ADC sampling rate | 25000 samples/s | Not specified | 25000 samples/s | Not specified | 25000 samples/s | Not specified |
| Status | Active | Active | Active | Active | Active | Active |
| Datasheet | PDF — IFD2410-1 (opens in a new tab) | PDF — IFD2410-1/IE (opens in a new tab) | PDF — IFD2410-3 (opens in a new tab) | PDF — IFD2410-3/IE (opens in a new tab) | PDF — IFD2410-6 (opens in a new tab) | PDF — IFD2410-6/IE (opens in a new tab) |
The industries we serve
Frequently asked questions
Answers to the most common questions about our products and services
No. Sensor and measuring controller share one housing without an external fiber. Appropriate power, data connections and installation accessories are still needed.
The table specifies 100 Hz to 8 kHz adjustment. Check resolution and averaging conditions separately rather than assuming static performance at maximum rate.
No. Its table specifies one layer, while IFD2415 supports multilayer measurement. Match the system to interface count and material thickness.
No. The 1, 3 and 6 mm spans specify 12, 18 and 24 µm spot diameters at mid-range. Spot size is separate from accuracy.
The detailed table specifies front IP64 for the sensor and IP65 for the controller section. Preserve this distinction instead of relying only on the housing headline.
Get a quote or advice