induSENSOR MSC controller components Inductive (eddy current)
induSENSOR MSC controller components: measurement controllers by Micro-Epsilon for inductive (eddy current).
Visit the Micro-Epsilon site (opens in a new tab)
Shared technical specifications for the series
Properties every model in the series has. What differs is in the models table below.
Range across models
The span covered by the models of the series together — not the range of each model.
| Number of measurement channels | 1–2 |
|---|
Models in the series
One column per model: what differs between the models of the series.
| Parameter | MSC7401 eli-m-450629a3f8114c1894547631 | MSC7602 eli-m-5eba84aa5c6c9d828434dd38 | MSC7802 eli-m-eb9a8145e0311983eef9d67f |
|---|---|---|---|
| Number of measurement channels | 1 | 2 | 2 |
| Interfaces | rs485 | rs485 | rs485 |
| Linearity (% FSO) | 0.02 % | 0.02 % | 0.02 % |
| Resolution (% FSO) | 0.012 % | 0.012 % | 0.012 % |
| Channel query-response time Maximum rate with a compatible controller | 2 ms | 2 ms | 2 ms |
| Status | Active | Active | Active |
| Datasheet | PDF — MSC7401 (opens in a new tab) | PDF — MSC7602 (opens in a new tab) | PDF — MSC7802 (opens in a new tab) |
The industries we serve
Frequently asked questions
Answers to the most common questions about our products and services
MSC7401 has one channel; MSC7802 and MSC7602 have two. MSC7602 supports bus chains, but unit count is not the number of inputs per unit.
Yes. The controller specifies linearity up to ±0.02% FSO separately from resolution. System selection must consider head, controller and their calibration.
No. The table distinguishes LVDT from half-bridge and 50 from 300 Hz. Bit resolution is not uniform across sensor types and settings.
No. MSC7401/MSC7802 specify IP67 with plugged connections; MSC7602 specifies IP20. Select for the required ingress protection.
Through the appropriate interface module according to the model table. A system-level protocol option does not mean it is onboard the controller.
Get a quote or advice
Tell us what you need and an engineer will come back to you.