Shared technical specifications for the series

Properties every model in the series has. What differs is in the models table below.

Range across models

The span covered by the models of the series together — not the range of each model.

Range across the models of scanCONTROL 2950
Accuracy 1–12 µm
X resolution 8.359375–112.109375 µm
X measuring range 10.7–143.5 mm
Working distance 52.5–390 mm
Z measuring range 8–265 mm

Models in the series

One column per model: what differs between the models of the series.

Models of scanCONTROL 2950 compared by parameter
Parameter scanCONTROL 2950-10 eli-m-ace47ad2b102e4f6de97c065 scanCONTROL 2950-100 eli-m-e68ed04fffb68a59db3033b3 scanCONTROL 2950-25 eli-m-11048579874d430925ce7409 scanCONTROL 2950-50 eli-m-8fcea16e8a111463380fcd77
Protection class (DIN EN 60529) ip65 ip65 ip65 ip65
Accuracy 1 µm 12 µm 2 µm 4 µm
interface_specification Digital I/O Digital I/O Digital I/O Digital I/O
Points per profile 1280 points/profile 1280 points/profile 1280 points/profile 1280 points/profile
X resolution 8.359375 µm 112.109375 µm 22.890625 µm 46.875 µm
Y measuring range 0 mm 0 mm 0 mm 0 mm
Interfaces profinet profinet profinet profinet
X measuring range 10.7 mm 143.5 mm 29.3 mm 60 mm
source_three_d_scanner_type Laser line Laser line Laser line Laser line
Working distance 52.5–60.5 mm 125–390 mm 53–79 mm 65–125 mm
Z measuring range 8 mm 265 mm 26 mm 60 mm
source_sampling_rate_basis scanCONTROL 2950-10 catalog maximum profile frequency scanCONTROL 2950-100 catalog maximum profile frequency scanCONTROL 2950-25 catalog maximum profile frequency scanCONTROL 2950-50 catalog maximum profile frequency
ADC sampling rate 2000 samples/s 2000 samples/s 2000 samples/s 2000 samples/s
quantity_specification X field up to 10.7 mm and Z range up to 8 mm; Y is generated by encoder-synchronized travel. X field up to 143.5 mm and Z range up to 265 mm; Y is generated by encoder-synchronized travel. X field up to 29.3 mm and Z range up to 26 mm; Y is generated by encoder-synchronized travel. X field up to 60 mm and Z range up to 60 mm; Y is generated by encoder-synchronized travel.
Operating temperature 0–45 °C 0–45 °C 0–45 °C 0–45 °C
Status Active Active Active Active
Datasheet PDF — scanCONTROL 2950-10 (opens in a new tab) PDF — scanCONTROL 2950-100 (opens in a new tab) PDF — scanCONTROL 2950-25 (opens in a new tab) PDF — scanCONTROL 2950-50 (opens in a new tab)

The industries we serve

Frequently asked questions

Answers to the most common questions about our products and services

It is HIGHSPEED. It provides calibrated profiles for external processing; do not assign another model’s SMART program.

Yes. Its documented speed class is up to 2 kHz, with 1280 points per profile. Application performance also depends on exposure and acquisition configuration.

No. Also match z interval, start distance, distance-dependent x width and model linearity. Extended range is distinguished from standard range.

Profiles are transferred over Ethernet/GigE Vision for suitable software. Plan evaluation/output for control values rather than assuming onboard SMART output.

Suffixes such as BL, SI, PT and VT identify different options. Use the exact model code; do not transfer another variant’s image, wiring or safety feature.

Didn't find the answer you were looking for?