Shared technical specifications for the series
Properties every model in the series has. What differs is in the models table below.
Range across models
The span covered by the models of the series together — not the range of each model.
| Measuring window | 0–2.4 mm |
|---|---|
| Resolution | 0.000075–0.024 µm |
| Measuring range | 0.5–1.2 mm |
| Linearity | 0.25–1.2 µm |
| Recommended target diameter | 5–17 mm |
| active_area_published_size | 2.9–8.1 mm |
| Weight | 0.03–0.038 kg |
| Number of measurement channels | 1–4 |
| Accuracy | 0.25–1.2 µm |
| ADC sampling rate | 1000–5000 samples/s |
Models in the series
One column per model: what differs between the models of the series.
| Parameter | CSH-0,25/SE/CAm-1,4 eli-m-11889d79f41cc307cf076bf6 | CSH-0,5/SE/CAm-1,4 eli-m-c4b344690032eb854afa2d2f | CSH-1,2/SE/CAm-1,4 eli-m-e60c8fada7a45930220f5c27 | CSH-1/SE/CAm-1,4 eli-m-11ca0fcbed289320d1aef2f2 | CSH-2/SE/CAm-1,4 eli-m-59d9e72560551d117bc7d3de |
|---|---|---|---|---|---|
| Start of measuring range | 0 mm | 0 mm | 0 mm | 0 mm | 0 mm |
| Operating temperature | -269–200 °C | -269–200 °C | -269–200 °C | -269–200 °C | -269–200 °C |
| source_recommended_mounting_position_mm | 3 mm | 3 mm | 3 mm | 3 mm | 3 mm |
| interface_specification | 4-20mA | Analog 0-10V | Analog 0-10V | Analog 0-10V | Analog 0-10V |
| Measuring window | 0–0.125 mm | 0–1 mm | 0–2.4 mm | 0–1 mm | 0–2 mm |
| Resolution | 7.499999999999999e-05 µm | 0.00015 µm | 0.024 µm | 0.0003 µm | 0.0005999999999999999 µm |
| Measuring range | 0.5 mm | 0.5 mm | 1.2 mm | 1 mm | 1 mm |
| Interfaces | ethernet | ethernet | ethernet | ethernet | ethernet |
| Linearity | 0.25 µm | 0.35 µm | 1.2 µm | 0.6 µm | 0.6 µm |
| Recommended target diameter | 5 mm | 7 mm | 11 mm | 11 mm | 17 mm |
| active_area_published_size | 2.9 mm | 4.1 mm | 6.3 mm | 5.7 mm | 8.1 mm |
| source_sampling_rate_basis | DT6120/IP: IP68 single-channel controller with 1 kHz analog bandwidth and up to 2 kSamples/s over RS485 | DL6222: 20 kHz bandwidth; DT6222 digital rate 3.906 kSamples/s | DL6228: 1 kHz bandwidth and 3.906 kSamples/s digital rate; optimized for CSE/HT sensors | DL6510 with CPM6011: 8.5 kHz bandwidth in the official maximum-eight-channel DT6530 base system | DT6110/IP: IP68 single-channel controller with 1 kHz bandwidth |
| Storage temperature | -269–200 °C | -269–200 °C | -269–200 °C | -269–200 °C | -269–200 °C |
| Weight | 0.03 kg | 0.03 kg | 0.033 kg | 0.033 kg | 0.038 kg |
| Protection class (DIN EN 60529) | ip40 | ip40 | ip40 | ip40 | ip40 |
| source_linearity_basis | Official sensor linearity from the current capaNCDT catalog; total measuring-channel linearity also includes the DT6530 (2-channel base) + DL6510 + CP6001 controller contribution. | Official sensor linearity from the current capaNCDT catalog; total measuring-channel linearity also includes the DT6230 + DL6230 controller contribution. | Official sensor linearity from the current capaNCDT catalog; total measuring-channel linearity also includes the DT6230 + DL6220 controller contribution. | Official sensor linearity from the current capaNCDT catalog; total measuring-channel linearity also includes the DT6530 (8-channel base) + DL6510 + CPM6011 controller contribution. | Official sensor linearity from the current capaNCDT catalog; total measuring-channel linearity also includes the DT6110/IP controller contribution. |
| quantity_specification | 0-0.25 mm model-level measuring interval | 0-0.5 mm model-level measuring interval | 0-1.2 mm model-level measuring interval | 0-1 mm model-level measuring interval | 0-2 mm model-level measuring interval |
| source_temperature_stability_um_per_k | -0.04 | -0.05 | -0.05 | -0.06 | -0.04 |
| Number of measurement channels | 1 | 1 | 4 | 4 | 2 |
| Contact | Non-contact | Non-contact | Non-contact | Non-contact | Non-contact |
| Accuracy | 0.25 µm | 0.35 µm | 1.2 µm | 0.6 µm | 0.6 µm |
| ADC sampling rate | 5000 samples/s | 5000 samples/s | 5000 samples/s | 5000 samples/s | 1000 samples/s |
| Status | Active | Active | Active | Active | Active |
| Datasheet | PDF — CSH-0,25/SE/CAm-1,4 (opens in a new tab) | PDF — CSH-0,5/SE/CAm-1,4 (opens in a new tab) | PDF — CSH-1,2/SE/CAm-1,4 (opens in a new tab) | PDF — CSH-1/SE/CAm-1,4 (opens in a new tab) | PDF — CSH-2/SE/CAm-1,4 (opens in a new tab) |
The industries we serve
Frequently asked questions
Answers to the most common questions about our products and services
These cylindrical titanium heads address very low temperatures, cleanrooms, UHV and strong magnetic fields. Select configuration limits from the exact specification.
No. It specifies an integral standard 1.4 m cable with a type B connector and still requires a suitable capaNCDT controller.
Linearity includes head and controller. Resolution references specified cable and bandwidth conditions, so assess the complete measurement chain.
No. Its table specifies IP40. Vacuum/material cleanliness does not substitute for dust and water ingress protection.
Use the recommended circumferential clamping position and account for heat transfer through surrounding material. Installation is part of the performance conditions.