eddyNCDT 3300 Inductive (eddy current)
eddyNCDT 3300: sensors and measurement devices by Micro-Epsilon for inductive (eddy current).
Download the datasheet (opens in a new tab) This series on the Micro-Epsilon site (opens in a new tab)
Shared technical specifications for the series
Properties every model in the series has. What differs is in the models table below.
Range across models
The span covered by the models of the series together — not the range of each model.
| Storage temperature | -40–150 °C |
|---|---|
| Operating temperature | -40–150 °C |
| Linearity | 0.8–160 µm |
| Resolution | 0.04–4 µm |
| Resolution (% FSO) | 0.005–0.2 % |
| Minimum target diameter | 6–420 mm |
| Start of measuring range | 0.04–8 mm |
| Measuring range | 0.4–80 mm |
| Accuracy | 0.8–160 µm |
| ADC sampling rate | 25000–100000 samples/s |
Models in the series
One column per model: what differs between the models of the series.
| Parameter | ES04 eli-m-3679d91f71835cdddc15b312 | ES08 eli-m-042ac8f6ca1f8648f0403f1b | ES1 eli-m-7da272a9de158ce2f7b823ac | ES2 eli-m-33a26bd598e77331fc21aaac | ES4 eli-m-7ad6b9eed85a97558604d915 | EU05 eli-m-6fd237d9cfcf030b23e31536 | EU1 eli-m-71b6dc3c4bd2c454181ac8bb | EU15 eli-m-0714b0b9863c251cac78306b | EU22 eli-m-da5a5515043a6cce0ae32a96 | EU3 eli-m-3e83fcca929bbfb7c5d84c7e | EU40 eli-m-924fbf5a8ae283185226d237 | EU6 eli-m-536ad775b0a31b2b562c07a7 | EU8 eli-m-0a47d9582a6075af770d2487 | EU80 eli-m-4c82a4976124db23f003a6c6 |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Storage temperature | -20–150 °C | -20–150 °C | -20–150 °C | -20–150 °C | -20–150 °C | -20–150 °C | -40–150 °C | -20–150 °C | -20–150 °C | -20–150 °C | -20–150 °C | -20–150 °C | -20–150 °C | -20–150 °C |
| interface_specification | RS422 / RS485 | Analog 0-10V | Analog 0-10V | RS422 / RS485 | RS422 / RS485 | 4-20mA | 4-20mA | Digital I/O | Analog 0-10V | RS422 / RS485 | RS422 / RS485 | 4-20mA | RS422 / RS485 | 4-20mA |
| Protection class (DIN EN 60529) | ip64 | ip64 | ip64 | ip64 | ip50 | ip64 | ip50 | ip64 | ip64 | ip64 | ip64 | ip64 | ip64 | ip64 |
| Operating temperature | 0–150 °C | 0–150 °C | 0–150 °C | -20–150 °C | 0–150 °C | 0–150 °C | -40–150 °C | 0–150 °C | 0–150 °C | -20–150 °C | 0–150 °C | -20–150 °C | 0–150 °C | 0–150 °C |
| Linearity | 0.8 µm | 1.6 µm | 2 µm | 4 µm | 8 µm | 1 µm | 2 µm | 30 µm | 44 µm | 6 µm | 80 µm | 12 µm | 16 µm | 160 µm |
| Resolution | 0.04 µm | 0.04 µm | 0.05 µm | 0.1 µm | 0.2 µm | 0.05 µm | 0.05 µm | 0.75 µm | 1.1 µm | 0.15 µm | 2 µm | 0.3 µm | 0.4 µm | 4 µm |
| quantity_specification | 0-0.4 mm model-level measuring interval | 0-0.8 mm model-level measuring interval | 0-1 mm model-level measuring interval | 0-2 mm model-level measuring interval | 0-4 mm model-level measuring interval | 0-0.4 mm model-level measuring interval | 0-1 mm model-level measuring interval | 0-15 mm model-level measuring interval | 0-22 mm model-level measuring interval | 0-3 mm model-level measuring interval | 0-40 mm model-level measuring interval | 0-6 mm model-level measuring interval | 0-8 mm model-level measuring interval | 0-80 mm model-level measuring interval |
| Resolution (% FSO) | 0.2 % | Not specified | Not specified | Not specified | Not specified | 0.005 % | Not specified | Not specified | Not specified | Not specified | Not specified | Not specified | Not specified | Not specified |
| Interfaces | ethernet | ethercat | ethercat | ethercat | ethercat | ethercat | ethercat | ethernet | ethernet | ethercat | ethernet | ethernet | ethercat | ethercat |
| Linearity (% FSO) | 0.2 % | Not specified | Not specified | Not specified | Not specified | 0.2 % | Not specified | Not specified | Not specified | Not specified | Not specified | Not specified | Not specified | Not specified |
| Minimum target diameter | 6 mm | 7.5 mm | 12 mm | 18 mm | 27 mm | 9 mm | 15 mm | 111 mm | 156 mm | 36 mm | 210 mm | 54 mm | 72 mm | 420 mm |
| Start of measuring range | 0.04 mm | 0.08 mm | 0.1 mm | 0.2 mm | 0.4 mm | 0.05 mm | 0.1 mm | 1.5 mm | 2.2 mm | 0.3 mm | 4 mm | 0.6 mm | 0.8 mm | 8 mm |
| Measuring range | 0.4 mm | 0.8 mm | 1 mm | 2 mm | 4 mm | 0.4 mm | 1 mm | 15 mm | 22 mm | 3 mm | 40 mm | 6 mm | 8 mm | 80 mm |
| source_sampling_rate_basis | DT3301: 100 kHz controller bandwidth; a separate acquisition-rate specification is not published | DT3301: 100 kHz controller bandwidth; a separate acquisition-rate specification is not published | DT3301: 100 kHz controller bandwidth; a separate acquisition-rate specification is not published | DT3300: 25 kHz controller bandwidth; a separate acquisition-rate specification is not published | DT3301: 25 kHz controller bandwidth; a separate acquisition-rate specification is not published | DT3301: 100 kHz controller bandwidth; a separate acquisition-rate specification is not published | DT3300: 100 kHz controller bandwidth; a separate acquisition-rate specification is not published | DT3301: 25 kHz controller bandwidth; a separate acquisition-rate specification is not published | DT3300: 25 kHz controller bandwidth; a separate acquisition-rate specification is not published | DT3300: 25 kHz controller bandwidth; a separate acquisition-rate specification is not published | DT3300: 25 kHz controller bandwidth; a separate acquisition-rate specification is not published | DT3301: 25 kHz controller bandwidth; a separate acquisition-rate specification is not published | DT3300: 25 kHz controller bandwidth; a separate acquisition-rate specification is not published | DT3300: 25 kHz controller bandwidth; a separate acquisition-rate specification is not published |
| source_sensor_mounting_specification | Mounting: Screw connection (table cell; model ES04) | Mounting: Screw connection (table cell; model ES08) | Mounting: Screw connection (table cell; model ES1) | Mounting: Screw connection (table cell; model ES2) | Mounting: through-holes (table cell; model ES4) | Mounting: Screw connection (table cell; model EU05) | Mounting: Screw connection (table cell; model EU1) | Mounting: through-holes (table cell; model EU15) | Mounting: through-holes (table cell; model EU22) | Mounting: Screw connection (table cell; model EU3) | Mounting: through-holes (table cell; model EU40) | Mounting: through-holes (table cell; model EU6) | Mounting: through-holes (table cell; model EU8) | Mounting: through-holes (table cell; model EU80) |
| Number of measurement channels | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 |
| Contact | Non-contact | Non-contact | Non-contact | Non-contact | Non-contact | Non-contact | Non-contact | Non-contact | Non-contact | Non-contact | Non-contact | Non-contact | Non-contact | Non-contact |
| Accuracy | 0.8 µm | 1.6 µm | 2 µm | 4 µm | 8 µm | 1 µm | 2 µm | 30 µm | 44 µm | 6 µm | 80 µm | 12 µm | 16 µm | 160 µm |
| ADC sampling rate | 100000 samples/s | 100000 samples/s | 100000 samples/s | 25000 samples/s | 25000 samples/s | 100000 samples/s | 100000 samples/s | 25000 samples/s | 25000 samples/s | 25000 samples/s | 25000 samples/s | 25000 samples/s | 25000 samples/s | 25000 samples/s |
| Status | Active | Active | Active | Active | Active | Active | Active | Active | Active | Active | Active | Active | Active | Active |
| Datasheet | PDF — ES04 (opens in a new tab) | PDF — ES08 (opens in a new tab) | PDF — ES1 (opens in a new tab) | PDF — ES2 (opens in a new tab) | PDF — ES4 (opens in a new tab) | PDF — EU05 (opens in a new tab) | PDF — EU1 (opens in a new tab) | PDF — EU15 (opens in a new tab) | PDF — EU22 (opens in a new tab) | PDF — EU3 (opens in a new tab) | PDF — EU40 (opens in a new tab) | PDF — EU6 (opens in a new tab) | PDF — EU8 (opens in a new tab) | PDF — EU80 (opens in a new tab) |
The industries we serve
Frequently asked questions
Answers to the most common questions about our products and services
The archived manufacturer manual describes 3300/3301. It supports identification of existing systems; new-purchase availability needs separate confirmation.
Sensor, sensor cable, adaptation board, controller, signal cable and power supply. The adaptation board participates in calibration, not just connection.
Not automatically. The manual requires adaptation-board changes and a calibration check when head type, cable length or target material changes.
No. It is specified for ranges up to 1 mm. Response and resolution depend on settings; static resolution is not the high-speed specification.
No. DT3300 uses ±12 V and 5.2 V; DT3301 specifies 11–32 VDC. Match power and synchronization cables as well.
Get a quote or advice
Tell us what you need and an engineer will come back to you.