IFS2405
IFS2405: sensors and measurement devices by Micro-Epsilon for distance & displacement sensors.
Visit the Micro-Epsilon site (opens in a new tab)
Shared technical specifications for the series
Properties every model in the series has. What differs is in the models table below.
Range across models
The span covered by the models of the series together — not the range of each model.
| Maximum measuring angle | 5–34 deg |
|---|---|
| Light spot diameter | 0.006–0.06 mm |
| Weight | 0.125–1.2 kg |
| Resolution | 0.004–0.747 µm |
| Number of measurement channels | 1–2 |
| Linearity | 0.09–5.6 µm |
| Measuring range | 0.3–30 mm |
| Start of measuring range | 6–220 mm |
| Minimum target thickness | 0.015–2.2 mm |
| ADC sampling rate | 8000–30000 samples/s |
| Accuracy | 0.09–9 µm |
Models in the series
One column per model: what differs between the models of the series.
| Parameter | IFS2405-0.3 eli-m-7e4c77cc22c2045d315c8a90 | IFS2405-1 eli-m-8f721e47221090ceb98747fb | IFS2405-10 eli-m-e3e4e6daa4294355af6602d4 | IFS2405-28 eli-m-6213bcdef17d699cd4c82f46 | IFS2405-28/VAC(001) eli-m-2c7559a27bdfa05d58fac324 | IFS2405-3 eli-m-03f370e9256ea6c314bf948f | IFS2405-30 eli-m-db464919880ce039e09839b0 | IFS2405-6 eli-m-c9c271ea10debf06132c1d58 | IFS2405/90-6 eli-m-65a16fed28e0c05dbb635a2c |
|---|---|---|---|---|---|---|---|---|---|
| source_numerical_aperture | 0.6 | 0.55 | 0.3 | 0.1 | 0.1 | 0.45 | 0.2 | 0.22 | 0.22 |
| Storage temperature | -20–70 °C | -20–70 °C | -20–70 °C | -20–70 °C | -20–70 °C | -20–70 °C | -20–70 °C | -20–70 °C | -20–70 °C |
| Maximum measuring angle | 34 deg | 30 deg | 17 deg | 5 deg | 5 deg | 24 deg | 9 deg | 10 deg | 10 deg |
| Light spot diameter | 0.006 mm | 0.008 mm | 0.016 mm | 0.06 mm | 0.06 mm | 0.009000000000000001 mm | 0.05 mm | 0.031 mm | 0.031 mm |
| Protection class (DIN EN 60529) | ip40 | ip64 | ip40 | ip64 | ip40 | ip40 | ip64 | ip64 | ip64 |
| Weight | 0.14 kg | 0.125 kg | 0.5 kg | 0.75 kg | 1.2 kg | 0.225 kg | 0.73 kg | 0.26 kg | 0.315 kg |
| Resolution | 0.018 µm | 0.004 µm | 0.018 µm | 0.06499999999999999 µm | 0.7469999999999999 µm | 0.008 µm | 0.53 µm | 0.017 µm | 0.19 µm |
| Interfaces | ethercat | profinet | ethercat | ethercat | Not specified | ethercat | ethercat | ethercat | ethernet |
| source_sampling_rate_basis | IFC2421: 10,000 measurements/s | IFC2421MP: 10,000 measurements/s | IFC2422MP: 10,000 measurements/s per synchronous channel | IFC2411: 8,000 measurements/s | Not specified | IFC2417: 25,000 measurements/s per synchronous channel | IFC2466: 30,000 measurements/s per synchronous channel | IFC2417: 25,000 measurements/s per synchronous channel | IFC2421: 10,000 measurements/s |
| quantity_specification | 0.3 mm net measuring range | 1 mm net measuring range | 10 mm net measuring range | 28 mm net measuring range | Not specified | 3 mm net measuring range | 30 mm net measuring range | 6 mm net measuring range | 6 mm net measuring range |
| Number of measurement channels | 1 | 2 | 1 | 2 | Not specified | 2 | 1 | 1 | 2 |
| Linearity | 0.09 µm | 0.2 µm | 1.5 µm | 5.6 µm | Not specified | 0.6 µm | 4.5 µm | 1.2 µm | 1.2 µm |
| Operating temperature | 5–70 °C | 5–70 °C | 5–70 °C | 5–70 °C | 5–70 °C | 5–70 °C | 5–70 °C | 5–70 °C | 5–70 °C |
| source_target_material | reflective, diffuse as well as transparent surfaces (e.g. glass) | reflective, diffuse as well as transparent surfaces (e.g. glass) | reflective, diffuse as well as transparent surfaces (e.g. glass) | reflective, diffuse as well as transparent surfaces (e.g. glass) | reflective, diffuse as well as transparent surfaces (e.g. glass) | reflective, diffuse as well as transparent surfaces (e.g. glass) | reflective, diffuse as well as transparent surfaces (e.g. glass) | reflective, diffuse as well as transparent surfaces (e.g. glass) | reflective, diffuse as well as transparent surfaces (e.g. glass) |
| source_linearity_basis | https://www.micro-epsilon.com/fileadmin/download/products/cat--confocalDT--en-us.pdf published optical-probe linearity; axial resolution is not substituted for linearity. | https://www.micro-epsilon.com/fileadmin/download/products/cat--confocalDT--en-us.pdf published optical-probe linearity; axial resolution is not substituted for linearity. | https://www.micro-epsilon.com/fileadmin/download/products/cat--confocalDT--en-us.pdf published optical-probe linearity; axial resolution is not substituted for linearity. | https://www.micro-epsilon.com/fileadmin/download/products/cat--confocalDT--en-us.pdf published optical-probe linearity; axial resolution is not substituted for linearity. | Not specified | https://www.micro-epsilon.com/fileadmin/download/products/cat--confocalDT--en-us.pdf published optical-probe linearity; axial resolution is not substituted for linearity. | https://www.micro-epsilon.com/fileadmin/download/products/cat--confocalDT--en-us.pdf published optical-probe linearity; axial resolution is not substituted for linearity. | https://www.micro-epsilon.com/fileadmin/download/products/cat--confocalDT--en-us.pdf published optical-probe linearity; axial resolution is not substituted for linearity. | https://www.micro-epsilon.com/fileadmin/download/products/cat--confocalDT--en-us.pdf published optical-probe linearity; axial resolution is not substituted for linearity. |
| Measuring range | 0.3 mm | 1 mm | 10 mm | 28 mm | 28 mm | 3 mm | 30 mm | 6 mm | 6 mm |
| source_mounting_description | Radial clamping (mounting adapter see accessories) | Radial clamping (mounting adapter see accessories) | Radial clamping (mounting adapter see accessories) | Radial clamping (mounting adapter see accessories) | Radial clamping (mounting adapter see accessories) | Radial clamping (mounting adapter see accessories) | Radial clamping (mounting adapter see accessories) | Radial clamping (mounting adapter see accessories) | Radial clamping (mounting adapter see accessories) |
| Start of measuring range | 6 mm | 10 mm | 50 mm | 220 mm | 220 mm | 20 mm | 100 mm | 63 mm | 41 mm |
| Minimum target thickness | 0.015 mm | 0.05 mm | 0.5 mm | 2.2 mm | 2.2 mm | 0.15 mm | 1.5 mm | 0.3 mm | 0.3 mm |
| interface_specification | RS422 / RS485 | RS422 / RS485 | RS422 / RS485 | Analog 0-10V | Not specified | RS422 / RS485 | Digital I/O | Digital I/O | Digital I/O |
| Contact | Non-contact | Non-contact | Non-contact | Non-contact | Non-contact | Non-contact | Non-contact | Non-contact | Non-contact |
| ADC sampling rate | 25000 samples/s | 25000 samples/s | 8000 samples/s | 30000 samples/s | Not specified | 10000 samples/s | 10000 samples/s | 30000 samples/s | 25000 samples/s |
| Accuracy | 0.09 µm | 0.3 µm | 3 µm | 8.4 µm | Not specified | 0.9 µm | 9 µm | 1.8 µm | 1.2 µm |
| Working distance | Not specified | Not specified | Not specified | Not specified | Not specified | 5–30 mm | Not specified | Not specified | Not specified |
| Status | Active | Active | Active | Active | Active | Active | Active | Active | Active |
The industries we serve
Frequently asked questions
Answers to the most common questions about our products and services
It is a universal confocal-head family for distance and one-sided transparent thickness, with short and long spans and radial-beam versions.
No. It depends on model and span, and accuracy decreases towards the angle limit. Select for the actual target slope.
The table includes IFS2405-28/VAC(001) with a specific housing, environment suitability and IP40. Vacuum suitability must not be assigned to every head in the family.
Not as specified. Static resolution uses 2048-value averaging at 1 kHz at mid-range. Dynamic resolution is specified separately as RMS noise.
Add a compatible fiber for its FC socket and a suitable confocalDT controller. Thickness selection also requires minimum thickness, refractive index and working range.
Get a quote or advice
Tell us what you need and an engineer will come back to you.